EasyManua.ls Logo

Commodore 1571 - Page 30

Commodore 1571
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1571/128D ALIGNMENT/REPAIR TEST
VERSION 1.2.1
+ — — — — — — — — — — — — -------------— — — — — — — — — —
-----------
— — — — — — — — — — — — — — -----------— — — —
----
— — — — — — — — +
| 1571/C128D ZERO STOP DIS-ASEMBLY AND SET-UP PROCEDURES |
+-------------------------------------------------------------------------------+
1. 1571 ONLY - Make Sure Alignment/Repair Main Menu is Displayed
2. 1571 ONLY - Turn Drive Power OFF
C128D ONLY - Turn System Power OFF
3. Disconnect Power Cable and Serial Cable (If Applicable)
4. 1571 ONLY - Remove the Four (4) Screws from the Bottom Case
and Remove the Top Case from the unit
C128D ONLY - Remove the Five (5) Screws from around the
Top Case and Remove Top Case from the unit
5. 1571 ONLY - Remove the Drive FacePlate and Set it Beside the Unit
6. Remove the Two (2) Shield Screws and Remove the Drive Shield
* The Drive Shield may not be installed in all units
7. Connect Power Cable and Serial Cable, (If Applicable)
8. 1571 ONLY - Turn Drive Power ON
9. C128D ONLY - Turn System Power ON, Load System Configuration,
Select C128D as SYSTEM TYPE, Load C128D Diagnostic
Menu, Load Alignment/Repair Test
10.
-
Select Zero Stop Test/Adjust Option from Diagnostic Menu
-
+
+--------------------------------------------------------------------------------------------------------------------+
| 1571/C128D ZERO STOP ADJUSTMENT PROCEDURES |
+--------------------------------------------------------------------------------------------------------------------+
| DO NOT ADJUST THE STOP UNTIL RADIAL ALIGNMENT IS CORRECT |
+-------------------------------------------------------------------------------+
1. Loosen the two (2) Stop Fixing Screws and rotate the Stop Plate
all the way towards the rear of the drive
* Press SPACE - Set Zero Stop
2. The Stepper Pulley will rotate to the Adjustment Position
3. NEWTRONICS - Slide the Stop forward and use a Standard Feeler
Gauge, to adjust the Stop to .006 - .010 of an inch
(.008 is Optimal)
4. NEWTRONICS - Tighten the two (2) Stop Fixing Screws
* The Head Must Not Move When The Screws Are Tightened
5. ALPS - Gently press down at the Top Center of the
Stop Plate until it stops
6. ALPS - Tighten the two (2) Stop Fixing Screws
(REAR SCREW FIRST)
* The Head Must Not Move When The Screws Are Tightened
7. Press SPACE - Re-Check Stop Adjust
* If the Zero Stop Test Fails ....
Repeat Steps 1 thru 4 until Adjustment is correct
* If the Zero Stop Test Passes ....
* Press RETURN - Return to Menu
Displays Alignment/Repair Test Main Menu
+-------------------------------------------------------------------------------+
Page 1-20

Related product manuals