S2VNA and S4VNA SCPI Programming Manual
171
DISP:WIND:ANN:MARK:SING
DISPlay:WINDow
<Ch>
:ANNotation:MARKer:SINGle[:STATe] {OFF|ON|0|1}
DISPlay:WINDow
<Ch>
:ANNotation:MARKer:SINGle[:STATe]?
Description
Selects display either the active trace markers or the all trace
markers.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
Choose from:
{ON|1} : Active trace markers
{OFF|0} : All trace markers
Query Response
{0|1}
Preset Value
1
Equivalent Softkeys
Markers > Properties > Active Only
DISP:WIND:MAX
DISPlay:WINDow
<Ch>
:MAXimize {OFF|ON|0|1}
DISPlay:WINDow
<Ch>
:MAXimize?
Description
Turn ON/OFF the active trace maximization inside the specified
channel.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Display > Active Trace/Channel > Maximize Trace

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish