EasyManua.ls Logo

Crystal AEOLUS IS-18CA - Page 28

Default Icon
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
26
8. Settings 8. Settings
0.0
0.0
0.0
0.0
90.0
70.0
50.0
25.0
d: Humidity Calibrate
Humidity can be calibrated by following the procedure below:
1. Set the instrument to the target humidity % (start with 25%)
2. Let instrument run and wait for the humidity % to stabilize
3. Measure humidity % inside chamber
4. Input the difference between set humidity % and actual humidity % with or
5. Repeat steps 1-4 for 50%, 70%, and 90%
0.0
0.0
0.0
0.0
1.0
3.0
5.0
10.0
c: CO2% Calibrate
CO2 % Concentration can be calibrated by following the procedure below:
1. Set the instrument to the target Co2% (start with 1%)
2. Let instrument run and wait for the CO2% to stabilize
3. Measure CO2% inside chamber
4. Input the difference between set CO2% and actual CO2% with “ ” or “ ”
5. Repeat steps 1-4 for 3%, 5%, and 10%