3.4.8
Relative
Signal
Strength
Measurements
Refer
to
Figure
3-10
for
the
following
test
procedure.
1)
Attach
antenna
to
RF
in/out
connector.For
measuring
remote
sites,
connect
to
an
outside
antenna.
2)
Select
RCV
mode.
3)
Program
frequency
to
be
monitored
in
MHz
using
the
numerical
keys
followed
by
the
ENTER
key.
4)
Open
Squench
control.
5)
Select
SHIFT
WATTS.
6)
Compare
signal
strength
reading
with
last
"known
good"
indication,
or
use
to
compare
two
transmitters
of
the
same
output
power,
and
approximate
location.
NOTE:
Several
comparisons
using
t~e
instrument
should
be
made
to
determine
what
a
"correct"
reading
for
a
local
transmitter
should
be.
This
may
also
require
taking
the
test
readings
at
the
same
time
of
day
as
the
sample
reading
due
to
changing
atmospheric
conditions.
7)
Signal
strength
meter
compression
begins
when
received
signals
exceed
3
mV
or
more.
Relative
signal
strength
can
be
meausred
above
this
level
by
inserting
an
external
pad
in
line
with
the
antenna
or
by
reducing
the
gain
of
the
antenna.
2500-MANUAL
PAGE
38