13. Troubleshooting
The Gauge will not Switch On
Are the batteries dead?
Check the batteries are inserted correctly.
Difficulty obtaining a Reading
If there is 1 single flashing bar on the display - this means the Gauge is
not receiving any echoes:
Check that the Probe-lead is properly connected to both Probe and
Gauge.
Check the condition of the lead, replace if necessary.
If there is mostly 1 fixed bar plus 1 flashing bar this means that the
Gauge is having difficulty obtaining more than one echo:
Check the Probe and its membrane are properly assembled.
If there are up to 3 fixed bars plus 1 flashing bar, but never any reading
- this means the Gauge is receiving unrelated echoes from more than
one reflector:
On heavily corroded areas this is often a problem, note this as a
problem area and try taking measurements in adjacent areas of
the same material.
Check the Gauge and Probe together on a test block, if there is
still no reading the Gauge may require servicing.
If Readings are Erratic or Unstable
Check that the Probe-lead is properly connected to both Probe and
Gauge.
Check that the Probe and its membrane are correctly assembled
with sufficient couplant between the probe face and membrane.
Check the Probe-frequency is suitable for the probable minimum
thickness of the material being measured. Probe frequencies too
low cause doubling and tripling of the actual thickness.
Check the gauge is not in Deep Coat mode.