M4-CYG3-M-ENG_Iss2.doc
Cygnus 3 Operating Manual
119
Tungsten Carbide
6660
0.2622
1.125
Epoxy Resin
2500
0.0986
0.422
Acrylic
2730
0.1076
0.461
Nylon (Polyamide)
2620
0.1032
0.443
Reading Conversions
If only a few measurements are to be taken on a material other
than Steel, it may be easier to leave the calibration set for Steel
and merely convert the readings by multiplying by the Conversion
Factor for the material being measured.
This method avoids unnecessary recalibration.
Example.
The Gauge is calibrated for Steel [5920 m/s], but the reading is
being taken on Copper [4700 m/s] :
T = t x V
COPPER
/ V
STEEL
= t x 4700 / 5920
= t x 0.794
thus : T = t x f [ where: f = V
COPPER
/ V
STEEL
]
where : T = true thickness of Copper being measured
t = actual reading obtained
f = Conversion Factor (from table)
V
COPPER
= Sound Velocity in Copper : 4700 m/s
V
STEEL
= Sound Velocity in Steel : 5920 m/s
The Conversion Factor f: is given for various materials
relative to steel in the Table of Sound Velocities on page 118
.