Cygnus 3 Operating Manual
39
Calibrating to a Known Thickness (Single Point)
This method of calibrating the Gauge is the most accurate as the
Gauge calculates the velocity of sound for the sample material.
1.
Accurately measure the
thickness of your sample
material.
2.
Place the Probe on the sample
so the Gauge is displaying a
thickness value.
3.
Press the MENU key to display
the MAIN MENU
4.
Scroll down to CALIBRATE then
press the OK key
5.
Use the +
-
keys to change
the thickness value as required,
shown in large numbers.
The Velocity of Sound is shown
at the bottom of the screen.
6.
Press the OK key to save the
calibration values and exit.
13.80
mm