What to do if cytiva WAVE Bioreactor 2/10 displays 'NEED 0% CAL or NEED 100% CAL'?
- DDustin BallAug 2, 2025
If your Cytiva Analytical Instruments displays 'NEED 0% CAL or NEED 100% CAL', it means the probe requires calibration. Try calibrating the probe. If the problem continues after calibration, the probe may need to be replaced.