DIMENSIONING SETUP AND CALIBRATION
125
5
If the length of the individual heads is within system tolerance but the length of the
MH BBOX FINAL is out of tolerance, adjust the X Shift in Modify Settings |
Dimensioning | Calibration by small increments. Run test boxes and make
adjustments until the length falls within system tolerance. If the length is too long,
reduce the absolute value of the X Shift slightly. If the length is too short, increase the
absolute value of the X Shift slightly.
If the width of the MH BBOX FINAL is out of tolerance, adjust the Y Shift for HEAD1 in
the Modify Settings | Dimensioning | Calibration (DC3000 User Interface) by small
increments. Run test boxes and make adjustments until the width falls within system
tolerance.
5. After the calibration process has been completed, save parameters from each DM3610
and the DC3000 Dimensioning Controller to the laptop computer. This is done using
Modify Settings | Save/Retrieve in DM3610 User Interface and DC3000 User Interface.
IMPORTANT: Be sure to turn Certification Mode back On
in both the DC3000 Dimensioning Controller and the
DM3610 dimensioning heads after calibration has been
completed in Legal for Trade applications.