6
1.4 AC
VOLTAGE
CALIBRATION
-
OPTION 10
ONLY
1.4.1
General
The procedure in the
table opposite is all that is
necessary to
completely 'AUTOCAL' the AC
voltage
function.
On
each range just a 'Zero', 'Gain' and
'AcHf'
calibration is
required.
If
the 'DVM Reading After
Calibration' is not in
accordance
with the
table,
repeat
operation
of the same
'CALIBRATE'
key is permissible to
improve
the
readings.
This will be necessary
with the
AcHf key.
1.4.2
Equipment
Required
A
copper
shorting
link and an AC
calibration
source
e.g.
Datron
4200
Autocal AC
Standard.
1.4.3
Checking Accuracy after
'AUTOCAL'
To
check the accuracy
after 'AUTOCAL' the
'Speci-
fication
Verification'
section
of the Operating
Handbook
will be
useful. It
provides
tables
for quick
reference
of
accuracy
on all ranges
and
functions
in
displayed
digits.