EasyManua.ls Logo

Datron 1062 - Page 16

Default Icon
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7B
AC
VOLTAGE
CALIBRATION
TABLE
(1061A
OPTION
12
ONLY)
Calibration
Operation
DC
coupled
AC
Zero
.IV Range
Zero
IV
Range
Zero
10V
Range
Zero
100V
Range
Zero
1000V
Range
Zero
10V
Full
Range LF
10V
Full
Range
HF
IV
Full
Range
LF
IV
Full
Range HF
.IV
Full
Range LF
.IV
Full
Range
HF
100V
Full
Range
LF
100V
Full
Range HF
1000V
LF
Range
Gain
I
1000V
HF
j
Range
Gain
Calibration
Source
Output
O.IOOmV
500Hz
(short
Guard
to Lo)
Short
Hi
to
Lo
to
Guard
0.00100V
500Hz
0.0100V
500Hz
0.100V
500Hz
1.00V
500Hz
10V
rms
500Hz
10V
rms
30KHZ
IV
rms
500Hz
IV
rms
30kHz
.IV rms
500Hz
.IV
rms
30kHz
100V
rms
500Hz
100V
rms
30kHz
500V
rms
500Hz
500V
rms
20kHz
DVM
Setting
AC,DC,.1
AC,
10
AC, 100
AC,
1000
AC,
10
AC, 10
AC,
100
AC, 100
'CALIBRATE'
Key
'Zero'
Check
only
'Zero'
'Zero'
'Gain'
'AcHf'
'Gain'
'AcHf'
'Gain'
'Gain'
'AcHf'
AC,1000
'KEYBOARD
500V'
'Gain'
DVM
Reading
After
Calibration
O.IOOmV
±10
digits
<C100 digits
0.00100V
±1
digit
0.010,0V
±1
digit
0.100
V
±1
digit
1.00V
±1 digit
10.000,0V
±1
digit
10.000.
0V
±10
digits
1
.000.
00V
±1
digit
1
.000,00V
±10
digits
1
00.000mV
±2
digits
1
00.000mV
±10
digits
100.000V
±1
digit
100.000V
±10
digits
500.00V
±1
digit
AC.1000
'KEYBOARD
500V'
500.00V
'AcHf'
±15
digits
Remarks
Set
'Local
Guard'.
Do
not
set
filter in.
Wait
for
reading
to
stabilize
before
operating
'Zero'
Lethal
voltage
present
-
increase
calibration
source
in
lOOv
steps
if
possible
Lethal
voltage
present
increase
calibration
source
in
100V
steps
if
possible. DO
NOT
EXCEED
25kHz