EasyManua.ls Logo

DePuy Mitek vapr vue User Manual

DePuy Mitek vapr vue
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Manufactured for DePuy Mitek by:
Gyrus Medical P/N: 192055-DA
VAPR
®
VUE
Wireless
Footswitches
P/N: 110008
Rev: F Revised 01/12
© 2010-2012 DePuy Mitek

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the DePuy Mitek vapr vue and is the answer not in the manual?

DePuy Mitek vapr vue Specifications

General IconGeneral
CategoryMedical Equipment
TypeElectrosurgical System
ManufacturerDePuy Mitek
ModelVAPR VUE
ApplicationArthroscopic Surgery
TechnologyRadiofrequency Ablation
PortabilityStationary
Intended UseSoft Tissue Ablation and Hemostasis
Control TypeFoot Pedal and Hand Control
Operating ModesCoagulation
Electrode TypeDisposable
Power SourceAC Power

Summary

Instructions for Use and Indications

Indications for Use

The VAPR VUE Wireless Footswitch is paired with the VAPR VUE Generator to activate the electrosurgical system.

Safety Warnings

Explosion Hazards

Do not use in flammable anesthetics; potential spark generation from battery operation.

Precautions for Operation

Operational and Maintenance Checks

Annual verification, daily inspection, and O-ring seal replacement are recommended for safety and performance.

Device Specifications and Usage

Operates on 2.405-2.480 GHz; use specified batteries and remove for long-term storage.

Operational Considerations

Signal Reception and Transmission

Details on RF activation, transmission direction, and pairing with generators.

Usage Scenarios and Storage

Advice on multiple units, tipping, backup activation, and proper storage for battery life.

Device Care and Operation

Cleaning Guidelines

IP68 rated, can be immersed. Use approved cleaners; do not autoclave or use ethylene oxide.

Environmental Handling

Allow device to reach room temperature before use if it is cold.

Electromagnetic Interference

Battery Replacement

Transmitter Battery Replacement Steps

Unscrew, remove cover, replace batteries observing polarity, and reassemble securely.

O-ring Seal and Battery Cover

Check and replace the O-ring seal if damaged; ensure the battery cover is flush.

Specifications

Environmental Conditions

Specifies transport, storage, and operating ambient temperature, humidity, and atmospheric pressure.