23
AdditionalInformation
MCLDriveControlUnit(seesystemoverview)
FortheCAREL/CARE/HOSPversionsthefollowingstandardswereapplied
BasedonEN60601-1:1990+A1:1993+A2:1995,
ElectromedicalEquipment
FortheCAREL/CAREversionthefollowingstandardswereapplied
BasedonEN1970:2000
BedsfortheDisabled
-
EN60601-1,mainsection2EnvironmentalConditions
EN60601-1,mainsection3ProtectionagainstElectricalShockHazard
EN60601-1,section21MechanicalStrength
EN60601-1,mainsection7Protectionagainstexcessivetemperatures
EN60601-1,mainsection9Operationoutsidetherequirementsandfaultscases
EN60601-1,mainsection10Constructionalrequirements
EN60601-1,section56.8Powersupplyindicator,optional
EN1970,inpartsection4-Unintentionalmovement:
Preventionviablockingdevice(e.g.lockingdevice
orMEDIXSKCAREorMEDIXSBCARE)or
IPROXX
Lowering ofthebackrestusingabatterybackupsystem
or
loweringofthebackrestusingamechanicalrelease
-Operator´scontrolwithkeypad
EN1970,section5.5.8Dimensionsofcontrolunit
EN1970,section5.6Operatingforcesofelectricalfunctions
En1970,section5.11Electrotechnicalrequirements
min.protectioncategoryIPX4
EN1970,section5.12Electromagneticcompatibility
®