b)Technicaldata
Powersupply.....................................................:230VAC50Hz
Powerconsumption..........................................:max.1.6AAC
Fuse..................................................................:T0.8A-T1.6A(accordingtothetrafotype)
Operatingmode.............................................:intermittentdutyID2min./18min.
Protectionclassification....................................:IorII
Currentinputofallslavedrives......................:max.8ADC(short-time)
Variations..........................................................:MCLCAREL/CARE/HOSP
Protectioncategory...........................................:IP44;IP66
Colours..............................................................:seesalesbrochure
Lengthxwidthxheightofthecontrolsystem...:175x139x90mm
Weight...............................................................:approx.2.2kg
Roomtemperature...........................................:from+10°to+40°C
Rel.humidity.....................................................:from 30%to75%
Barometricpressure.........................................:From700hPato1060hPa
1)
2)
Dimensionsandweights
Ambientandstorageconditions
1)Operatingmode=,i.e.runfor2min.max.underratedload,thenarest
periodof18min.mustbeobserved,otherwiseoperational!
2)Accordingtoregulations,nomorethantwodrivesmayberunsimultaneouslyattheratedload!
intermittentdutyID2min./18min.
failurecouldoccur
(Specificationofthecountries)