112
113
Electromagneticimmunity
The COM-DEX is intended for use in the electromagnetic environment
specified below. The customer or the user of a COM-DEX should assure
that it is used in such an environment.
Immunity Test IEC 60601
Test level
Compliance
level
Electromagnetic
environment
guidance
Electro-static
discharge
(ESD)
IEC
61000-4-2
± 6 kV con-
tact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humid-
ity should be at
least 30 %.