Declaration – Electromagnetic Emissions
Declaration – Electromagnetic Immunity
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity (IEC 60601-1-2)
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8kV contact
± 15kV air
± 8kV contact
± 15kV air
The relative humidity should be
at least 5%.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV For mains
supply lines
100kHz repetition
frequency
± 2 kV For mains supply
lines
100 kHz repetition
frequency
Mains supply quality for the
mains adapter should be that
of a typical commercial and/or
hospital environment.
Surge
IEC61000-4-5
± 2kV Line(s) to
ground
± 1kV line(s) to line
± 2kV Line(s) to ground
± 1kV line(s) to line
Mains supply quality for the
mains adapter should be that
of a typical commercial and/or
hospital environment.
Voltage dips, short interruptions
and voltage variations on mains
supply IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Mains supply quality for the
mains adapter should be
that of a typical commercial
and/or hospital environment.
In the event of a mains
interruption the VENTURI
system will automatically use
internal battery power,
unless the battery is exhausted.
0 % U
T
; 1
cycle
70 % U
T
; 25/30
cycles
Single phase: at 0°
0 % U
T
; 1 cycle
70 % U
T
; 25/30 cycles
Single phase: at 0°
Voltage interruptions 0 % U
T
; 250/300
cycle
0 % U
T
; 250/300 cycle
Mains frequency (50/60Hz)
magnetic field IEC61000-4-8
30 A/m 30 A/m Mains frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
and/or hospital environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions (IEC 60601-1-2)
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Class B
The VENTURI systems are suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonics emissions
61000-3-2
Class A
Voltage fluctuations / flicker emissions
61000-3-3
Complies