Product Specications
Temperature Ranges: User Selectable: -20 to +120˚F (-20 to +50˚C),
+40 to +110˚F (+5 to +40˚C)
Temperature Accuracy: ±1.8˚F (±1.0˚C)
Temperature Sensor: Thermistor
Humidity Range: 0 to 95% (non-condensing)
Humidity Accuracy: ±2% from 0 and 60%, ±3% from 60 to 95%
Humidity Sensor: Monolithic IC humidity sensor
Dew Point Range: -22 to +122˚F (-30 to +50˚C)
Dew Point Accuracy: 95% to 30% RH ±2˚C, ±3.6˚F
30% to 10% RH ±4˚C, ±7.2˚F
10% to 5% RH ±8˚C, ±14.4˚F
5% to 0% RH ±49˚C, ±88.2˚F
Ambient Operating Conditions: 32 to +122˚F (0 to 50˚C), 0 to 95% RH (non-condensing)
Display Resolution: LCD, 1˚F (1˚C), 1% RH
Chart Size: 8” (20.3 cm) DIA
Recording Time: User selectable: 24 hour, 7 day or 31 day
Average Response Time: 2.5 min. for 63% of full scale
NOTE: Response time is slower when using battery
Calibration: User calibration of all variables
Product Features
• Remote sensing capability with optional extension cords (10’, 50’ and 100’ available–see Accessories)
• Alarms: Audio/Visual with High/Low setting capabilities
• Microprocessor control and electronic sensing
• C/F and RH/Dew Point switchable
• Large 8” diameter circular chart
• Min/Max feature allows display of minimum and maximum readings
• User calibration
• Locking control keypad and door
• Large digital display shows two variables at once
• User selectable recording times and temperature ranges