Apendix
Error messages may eventually open on your unit screen and are
informational only. If one of these messages opens on your display
follow the instructions described below and if the problem persists
please send us a detailed report at
www.demeq.com/form_Support.html
Figure A.2: System error message
Error 1
Internal Error
Cause
Internal Error
Solutions
Shutdown the unit, wait a few seconds, and power
back on.
Contact dmq.
Error 2
Attempt to store a value over an existing value.
Cause
Improper unit shutdown (Example: Removing
batteries) and powering the unit back on to store
values in the Data Logger.
Solution
Download Data Logger values to PC or printer and
erase memory.
If a message with a different number where to appear please contact
dmq.