Pre-Installation Installation Training
A70622-D1
109/192
If the PLd is chosen, inputs 1,2,3,4 in the I/O window are automatically initiated and set
to their specific status.
- UI will force the In 1 label to “External Interlock 2”.
- UI will force the In 2 label to “Source safety loop”.
- UI will force the In 3 label to “Shutter safety loop”.
- UI will force the In 4 label to “Non usable”.
The inputs of the industrial board are automatically configured as follow :
Select the safety level in the Advanced settings
Note : Signals from Extractor or Air cooler (warning/Alarm) should be set from
inputs In5 if PLd option is active.
Step 8 : Configure the level in the software