Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 177
Configuration
11.3.6.2 Vertical tab "Units"
11.3.6.3 Vertical tab "System test"
Configuration
password
Change password Changes the configuration password.
Reset all pages to Factory defaults Resets all the settings on all pages in
the System setup dialog to the factory
defaults.
Headline/
Parameter
Setting range
Description
Weight kg; lbs Sets the units.
Height cm; in
Airway pressure mbar; hPa; cmH
2O
Supply pressure bar; kPa×100; psi
CO
2 %; kPa; mmHg
Volatile agents %; kPa
Headline/
Parameter
Setting range
Description
General
Always use walk-
through mode
On; Off Sets the walk-through mode function.
When On is configured, system tests
and leakage tests will always be exe-
cuted in walk-through mode.
Sample line is
connected during
test
On; Off Specifies whether the sample line is
connected to the Y-piece or to the filter
on the Y-piece during the automatic
tests.
When On is configured, the device can
automatically check whether, e.g.,
breathing hoses or central supply
hoses are incorrectly connected.
When Off is configured, additional
checks are required during the system
test.
Test gas supply
Headline/
Parameter
Setting range
Description