300 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
17.2.3 Ventilation modes and effective parameters
Group Tabs Ventilation mode
Basic parameter
(normal therapy
bar)
Additional parame-
ters
(expanded therapy
bar)
Manual ventila-
tion/spontaneous
breathing
Man/Spon Manual / Sponta-
neous
CPAP
1)
Assisted ventila-
tion
PSV
1)
CPAP / PSV Trigger
Psupp
RRmin
PEEP
Slope
Insp term
Pressure-con-
trolled ventilation
PC PC - CMV Pinsp
RR
PEEP
Ti
CMV
Slope
PC - SIMV Pinsp
Psupp
1)
= Off
RR
PEEP
Ti
SIMV
Trigger
Slope
PC - SIMV / PS
1)
Pinsp
Psupp
1)
>0
RR
PEEP
Ti
SIMV
Trigger
Insp term
1)
Slope

Table of Contents