Supplement | Atlan A300, A350 ceiling-mounted versions and wall-mounted version 55
Ceiling-mounted version with integrated castors
6.2.3 Docking to the ceiling supply unit
The following describes docking to a ceiling supply unit with lifting function, e.g.,
Movita lift.
Prerequisite:
If the device is docked with a ceiling supply unit with a key switch, the key switch
must be in the horizontal position.
1. Connect all necessary cables and hoses at the rear of the device:
Central supply hoses (1)
Power cable (2)
2. Check that the floor below the ceiling supply unit is level and cleared. There
must be no objects below the ceiling supply unit.