Instructions for use Babyleo TN500 SW 1.0n 159
Troubleshooting
! 217 Prewarming in progress The device is warming up. Wait until prewarming is fin-
ished.
To deactivate prewarming,
switch to manual mode.
! 215 Rotary knob faulty The rotary knob has been
pressed for more than 10
seconds.
Release the rotary knob.
Contact specialized service
personnel.
! 117 Scale calibration required An internal fault occurred. Disconnect the scale cable.
Contact specialized service
personnel.
! 115 Scale communication error A scale communication error
occurred.
Make sure that the scale
cable is correctly connected.
An internal fault occurred. Disconnect the scale cable.
Contact specialized service
personnel.
! 116 Scale disconnected The scale cable is not con-
nected.
Make sure that the scale
cable is correctly connected.
! 114 Scale faulty An internal fault occurred. Disconnect the scale cable.
Contact specialized service
personnel.
! 211 Secondary speaker faulty An internal fault of the sec-
ondary acoustic alarm sys-
tem occurred.
Contact specialized service
personnel.
!! 100 Skin mode not confirmed The mode was selected, but
it was not confirmed.
Select skin mode and press
the rotary knob to confirm.
!! 117 Skin temp. sensor faulty,
heating off
At least one of the skin tem-
perature sensors is faulty.
Replace the faulty skin tem-
perature sensor.
Contact specialized service
personnel.
!! 19 Skin temperature differ-
ence high
The current temperature dif-
ference between central skin
temperature (yellow) and
peripheral skin temperature
(white) is above the set
upper alarm limit.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
!! 18 Skin temperature differ-
ence low
The current temperature dif-
ference between central skin
temperature (yellow) and
peripheral skin temperature
(white) is below the set lower
alarm limit.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
!! 213 Skin temperature high The tolerated deviation for
central skin temperature (yel-
low) has been exceeded.
The current central skin tem-
perature is too high.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Alarm pri-
ority
Alarm Cause Remedy