Assembly and preparation
5.4 Establishing potential equalization
Differences in electrical potential between devices can be reduced by potential
equalization.
Potential equalization does not replace the protective ground connection.
5.4.1 Connecting the potential equalization cable
1 Connect the potential equalization cable to the potential equalization pin on the
device.
2 Connect the potential equalization cable to the hospital's potential equalization
connection (e.g., wall, ceiling supply unit, operating table).
5.5 MEDIBUS connection for data transfer
This chapter describes the software protocol for data transfer between the device
and an external medical or non-medical device.
MEDIBUS.X is a software protocol for data transfer between the device and an
external medical or non-medical device (e.g., hemodynamic monitors, data
management systems, or computers) via a COM port.
Note the following software protocols before transmitting any data:
MEDIBUS.X, Rules and Standards for Implementation (9052607)
MEDIBUS.X, Profile Definition for Data Communication V1.n (9052608)
Prerequisite:
The MEDIBUS cable is ready to use.
WARNING
Risk of electric shock
If the interface connectors and the patient are touched at the same time, there is a
risk of an electric shock.
Do not simultaneously touch the COM port and the patient.
CAUTION
Risk to patient due to incorrect data transfer
All data transferred via the COM port are for information only and must not be used
as the sole basis for diagnostic or therapeutic decisions. This may put the patient at
risk.
Do not use data transferred via the COM port as the sole source of information.
Procedure:
1 Connect an external device to the COM port.
2 Configure the COM port.