Configuration
102 Instructions for use Babylog 8000 plus SW 5.n
The following settings can be selected:
Ventilation parameters Scale range Measuring signal
range
Airway pressure –10 ... 90 mbar (–10 ... 90 cmH
2O) → 0 ... 10 V
–5 ... 45 mbar (–5 ... 45 cmH
2O) → 0 ... 10 V
Mean airway pressure –10 ... 90 mbar (–10 ... 90 cmH
2O) → 0 ... 10 V
–5 ... 45 mbar (–5 ... 45 cmH
2O) → 0 ... 10 V
FiO
2 0 ... 100 Vol.% → 0 ... 10 V
Flow –40 ... 40 L/min → 0 ... 10 V
–20 ... 20 L/min → 0 ... 10 V
–10 ... 10 L/min → 0 ... 10 V
–5 ... 5 L/min → 0 ... 10 V
Volume 0 ... 500 mL → 0 ... 10 V
0 ... 100 mL → 0 ... 10 V
0 ... 50 mL → 0 ... 10 V
0 ... 25 mL → 0 ... 10 V
Tidal volume 0 ... 500 mL → 0 ... 10 V
0 ... 100 mL → 0 ... 10 V
0 ... 50 mL → 0 ... 10 V
0 ... 25 mL → 0 ... 10 V
Minute volume 0 ... 10 L/min → 0 ... 10 V
0 ... 5 L/min → 0 ... 10 V
0 ... 1 L/min → 0 ... 10 V
0 ... 0.5 L/min → 0 ... 10 V
MVim 0 ... 10 L/min → 0 ... 10 V
0 ... 5 L/min → 0 ... 10 V
0 ... 1 L/min → 0 ... 10 V
0 ... 0.5 L/min → 0 ... 10 V
VTim 0 ... 500 mL → 0 ... 10 V
0 ... 100 mL → 0 ... 10 V
0 ... 50 mL → 0 ... 10 V
0 ... 25 mL → 0 ... 10 V
VThf 0 ... 25 mL → 0 ... 10 V
0 ... 5 mL → 0 ... 10 V