Instructions for use Babylog 8000 plus SW 5.n 103
Configuration
Setting the default values during operation
Press the key.
The default values are used.
If a measured value exceeds the limits of the
scale, the voltage is limited to the maximum value
on the scale.
Configuring the impulse output
The following signals are available:
– Triggered mandatory breath
– Mandatory breath
–Alarm
Default value: Mandatory breath
1 Open the dialog window by pressing the Cal.
Config. > Config > Com keys.
2 Press the Param key repeatedly until is
highlighted.
3 Use the or key to select the signal.
DCO
2 0 ... 200 mL
2
/s → 0 ... 10 V
0 ... 50 mL
2
/s → 0 ... 10 V
Continuous flow (set value) 0 ... 125 L/min → 0 ... 10 V
Leakage rate 0 ... 100 % → 0 ... 10 V
Spontaneous portion of minute vol-
ume
0 ... 100 % → 0 ... 10 V
Factory-set default values:
Analog1: Flow –20 ... 20 L/min
Analog2: Airway pressure –10 ... 90 mbar (–10 ... 90 cmH
2O)
Ventilation parameters Scale range Measuring signal
range
21153
21154