Configuration
104 Instructions for use Babylog 8000 plus SW 5.n
Configuring the RS232 interface
The following settings can be made:
– Transfer speed Baudrate
– Parity test Parity
Factory-set default values:
1 Open the dialog window by pressing the Cal.
Config. > Config > Com keys.
2 Press the Param key repeatedly until
Baudrate is highlighted.
3 Use the or key to select one of the
following values:
–9600
–2400
–1200
4 Press the Param key repeatedly until Parity is
highlighted.
5 Use the or key to select one of the
following values:
– NONE
– EVEN
– ODD
When using the printer, select NONE.
Configuring the printer
The printer must be configured as follows:
Baudrate 9600
Parity NONE (no parity test)
Stop bit 1 (fixed)
Data bit 8 (fixed)
21156
Baud rate As for Babylog 8000
Parity None
Data bit 8
Handshake
mode
XON/XOFF
Alternate Control Sequence
Mode