EasyManua.ls Logo

Dräger Babylog 8000 plus - Reprocessing Methods and Lists

Dräger Babylog 8000 plus
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleaning, disinfection and sterilization
114 Instructions for use Babylog 8000 plus SW 5.n
Reprocessing methods
Classification of medical devices
For reprocessing, the medical devices and their
components are classified according to their type
of application and the resulting risks:
Non-critical medical devices: Surfaces
accessible to the user and patient, e.g., device
surfaces, cables
Semi-critical medical devices: parts conducting
breathing gas, e.g., breathing hoses, masks
Testing of methods and agents
Cleaning, disinfection, and sterilization of medical
devices has been tested with the following
procedures and agents. At the time of testing, the
following procedures and agents showed good
material compatibility and effectiveness:
Non-critical medical devices
Manual disinfection and simultaneous cleaning:
Buraton 10F by Schülke & Mayr
Concentration: 0.5 %, contact time: 60 min
Concentration: 1 %, contact time: 30 min
Semi-critical medical devices
Manual cleaning:
Neodisher LM2 by Dr. Weigert
Sekusept powder classic by ECOLAB (Y-piece
flow sensor)
Manual disinfection:
Korsolex extra by Bode Chemie
Concentration: 3 %, contact time: 15 min
Machine cleaning:
Neodisher MediClean by Dr. Weigert
Machine disinfection:
Thermal, 93 °C (199.4 °F) for 10 min
Sterilization:
Hot steam, 134 °C (273.2 °F) for 5 min
Non-critical medical devices
Manual disinfection with simultaneous
cleaning
When selecting a suitable disinfectant, adhere to
the country-specific lists of disinfectants. The list of
the German Association for Applied Hygiene
(Verbund für Angewandte Hygiene VAH) applies in
German-speaking countries.
Strictly observe the manufacturer's specifications
on the disinfectants. Manufacturers may change
the composition of disinfectants over time.
Procedures:
1 Remove soiling immediately with a cloth
soaked in disinfectant.
2 Perform surface disinfection by scrubbing and
wiping.
3 Remove disinfectant residues after the contact
time has elapsed.
WARNING
Risk of electric shock or device malfunction
Liquid that enters into the device can cause
the device to malfunction or may damage the
device and endanger the patient.
Only scrub-and-wipe-disinfect device
surfaces and cables and make sure no liquids
penetrate into the device.

Other manuals for Dräger Babylog 8000 plus

Related product manuals