Instructions for use Babylog 8000 plus SW 5.n 123
Maintenance
– Correct functioning of the pneumatic safety
valve:
Before the check, switch off the device and
unplug the mains plug.
Pressure: 90 to 110 mbar (or hPa or
cmH
2O)
5 Check the electrical safety according to
IEC 62353. If a breathing gas humidifier or
power socket strip (e.g., on the trolley) are
used, they must be subjected to the same
check. The check must be performed on
individual devices and in the integrated system.
– Protective ground contact resistance
<0.3
– Replacement device leakage current
<1 mA
– Replacement patient leakage current
<5 mA
6 Perform a functional test of the following
features according to the instructions for use:
– All functions described in the test steps of
the device check.
– Functioning of the power supply failure
alarm
22610