Instructions for use Babylog 8000 plus SW 5.n 131
Technical data
Ambient conditions
Set values
During operation
Temperature 10 to 40 °C
(50 to 104 °F)
Pressure range 780 to 1060 hPa
(11.3 to 15.4 psi)
Relative humidity 30 % to 90 %, without condensation
During storage and transportation
Temperature –20 °C to 60 °C
(–4 °F to 140 °F)
Pressure range 500 to 1060 hPa
(7.3 to 15.4 psi)
Relative humidity 10 % to 95 %, without condensation
Depending on the accessories used, more stringent
ambient conditions can apply. Strictly observe the
corresponding instructions for use.
Unless explicitly stated, the tolerances in the techni-
cal data do not include the uncertainty of measure-
ment of external testing devices. The uncertainties
of measurement of the testing devices used are
available upon request.
Ventilation modes IPPV/IMV, IPPV, IMV
SIMV, SIPPV
PSV, CPAP
Inspiratory time TI
Range 0.1 to 2 s
Resolution 0.1 to 1 s: 0.01 s
1 to 2 s: 0.1 s
Accuracy ±10 ms