Technical data
132 Instructions for use Babylog 8000 plus SW 5.n
Expiratory time TE
Range 0.2 to 30 s
Resolution 0.2 to 1 s: 0.01 s
1 to 10 s: 0.1 s
10 to 30 s: 1 s
Accuracy ±10 ms
Tidal volume Vtset
Range 2 to 100 mL
Resolution 2 to 9.9 mL: 0.1 mL
10 to 19.5 mL: 0.5 mL
20 to 100 mL: 1 mL
Accuracy 2 to 5 mL: ±0.5 mL
>5 mL: ±10 %
Inspiratory flow
Range 1 to 30 L/min
Resolution 1 to 10 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
Accuracy ±10 %
Expiratory flow
Range 1 to 30 L/min
Resolution 1 to 10 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
Accuracy ±10 %
Inspiratory O
2 concentration FiO2
Range 21 to 100 Vol.%
Resolution 1 Vol.%
Accuracy ±3 %
Inspiratory pressure Pinsp
Range 5 to 80 mbar (5 to 80 cmH
2O)
Resolution 1 mbar (1 cmH
2O)
Accuracy ±1 mbar (±1 cmH
2O) or 3 % of the measured value,
whichever is greater