6173.3 Babylog 8000/8000 plus/8000 SC 12/00 General Page 10
For internal use only. Copyright reserved.
GBK61733XXTL1.fm 07.07.05
Babylog 8000/Babylog 8000 plus/Babylog 8000 SC Test Performance:
Fig. 5: Test Performanc e
Important: If during Test 1 the variant A and during Test 2 the variant B (pressure
values possibly too high) is recognized, then it is a B variant with a non-permissible
spring mounted in the valve Y 3.3. This spring must be removed and the test be
repeated. Mount the bottom cover.
Slowly compress the bellows and generate the following pressure:
Variant A : 25 mbar ±5 mbar.
Variant B : 100 mbar ±10 mbar.
1.5 bar
/21.3 psi
1.5 bar
/21.3 psi
Variant A
(not in the USA)
Variant B
Inspiratory block
Inspiratory block
Y 3.3
Y 3.3
25 mbar /
cmH
2
O
100 mbar /
cmH2O