12 Testing the Pressure Regulator
• Connect the inspiratory socket with the expiratory socket.
• Adjust the device as shown in the Figure below
Fig. 7: Pressure Regulator
• Connect the device to the O
2
supply only.
The following message is displayed on the device: “no medical air”.
• Connect the device to the medical AIR supply only.
The following message is displayed on the device: “no oxygen”.
30 15
CPAP
Flow sensor cable