6173.3 Babylog 8000/8000 plus/8000 SC 12/00 General Page 13
For internal use only. Copyright reserved.
GBK61733XXTL1.fm 07.07.05
12 Testing the Pressure Regulator
Connect the inspiratory socket with the expiratory socket.
Adjust the device as shown in the Figure below
Fig. 7: Pressure Regulator
Connect the device to the O
2
supply only.
The following message is displayed on the device: “no medical air”.
Connect the device to the medical AIR supply only.
The following message is displayed on the device: “no oxygen”.
30 15
CPAP
Flow sensor cable