All rights reserved. Copyright reserved.
_ _Printed on_18.05.05_F61733XXT01.fm
50
Dräger Medical AG & Co. KGaA 6173.3
Function description Babylog 8000
10.12 Potentiometer Field
The Potentiometer Field comprises the following components:
6 potentiometers (O
2
vol.%, T
I
, T
E
, Insp. Flow , P
insp
, PEEP/CPAP)
9 LEDs (O
2
vol.%, T
I
, T
E
, Insp. Flow , P
insp
, PEEP/CPAP, CPAP, IPPV, reserve)
3 keys (CPAP, IPPV, reserve) (Vent. Mode, Vent. Option, reserve as of software version 5.n
10.13 Display Field
The display field contains the following components:
11 keys (silence, reset, man./insp., menu, keys 1 to 6, light). The "light" key is only available on
Babylog 8000 with LC display. The key signals are read in via the Front Controller PCB or the Front
PCB.
LED array of the bargraph. The LED array is controlled via the Display PCB.
7 LEDs (warning (yellow), mode, monitoring, silence, triggers 1 to 3). The LEDs are controlled via the
Front Controller PCB or the Front PCB.
2 red alarm LEDs. The LEDs are controlled via the Monitoring PCB/Watchdog PCB.