Replacing Non-Repairable Items Babylog 8000
5 Replacing the NiCd Battery (Power Failure Alarm) Every 2 Years
Remove the screw 1 on the fan/loudspeaker cover.
Fig. 6: Opening the Loudspeaker Cover
Gently lower the cover 3.
Disconnect the cable connection 2 and place the cover 3 aside.
Remove the screws 4 on the battery mount.
Fig. 7: Removing the Loudspeaker Cover
1
2
3
4
4