Dräger Medical AG & Co. KGaA 6173.3
All rights reserved. Copyright reserved.
_ _Printed on_18.05.05_GBK61733XXV01.fm
67
Replacing Non-Repairable Items Babylog 8000
6 Replacing the Pressure Reducer Every 6 Years
• Disconnect the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC from the mains supply.
• Disconnect the gas supply hoses from the central gas supply/gas cylinder.
• Disengage the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC from the trolley. To do so, loosen
the knurled nuts 1 on the rear panel.
Fig. 9: Location of the Knurled Nuts
Fig. 10: Disengaging the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC
1 1