Babylog 8000 Error List
2 List of device error messages
If an error occurs, read the Babylog 8000/8000 SC error messages
on the 7-segment display of the Monitoring PCB!
Table 1: Device error messages
Display error
number
Monitoring PCB
error number
Meaning
000 CPU PCB ROM error
001 CPU PCB RAM error
002 One or more keys jammed
(continuously pressed).
003 Data transmission between
CPU PCB and the dual-port RAM
of the Monitoring PCB is
interfered.
004 Watchdog error or CPU PCB reset
005 CPU PCB error: The CPU PCB is
in the boot phase although the
Monitoring PCB is running in the
background (e.g. by a CPU PCB
reset).
006 Vector error or CPU PCB reset
500 Software error: Incorrect operating
mode, for example, error in
storage of operational data in the
time-keeper RAM on the CPU
PCB.
501 Monitoring PCB does not function
correctly
502 Flow PCB does not function
correctly
503 CPU PCB software error:
Impermissible breathing phase
504 CPU PCB software error