Dräger Medical AG & Co. KGaA
All rights reserved. Copyright reserved.
_ _Printed on_18.05.05_GB_K61733XXFL1.fm
93
Error List Babylog 8000
Display error
number
Monitoring PCB
error number
Meaning
505 CPU PCB software error
506 to 513 CPU PCB software error.
800 FiO
2
too high by more than 10
vol.%.
801 FiO
2
too low by more than 10
vol.%.
802 FiO
2
measurement deviation
between master (CPU PCB and
I/O PCB) and Monitoring PCB is
too high.
803 Measured Pinsp to potentiometer
setting Pinsp too high by more
than 25 mbar (cmH
2
O).
804 Pinsp greater than 110 mbar
(cmH2O).
805 Pinsp less than -5 mbar (cmH2O)
806 Flow measurement on the Flow
PCB is faulty
807 Potentiometer evaluation is faulty:
Master (CPU PCB and
Frontcontroller PCB) and
Monitoring PCB read different
values.
809 Battery or battery charge faulty
(battery voltage < 7 V)
Table 1: Device error messages