22 Dräger DrugTest 5000 Test System
Troubleshooting
10811
(during
measur
ement)
Low Battery warning Charge battery
Rechargeable battery is completely
depleted.
Connect mains adaptor or finish
measurement before device will be
switched off.
10811
(before
measur
ement)
Low Battery warning
Charge battery
Rechargeable battery is completely
depleted.
Connect mains adaptor or switch off
device.
20000
User error: Door is still open
and should be closed
Close door
The Analyzer's sliding door is still open.
Close door.
20001
User error: Cartridge and/or
cassette not removed
Remove cassette and close door
There is still a cassette or cartridge
in the device.
Remove cassette and cartridge and close
the door.
20002
User error: Entered date is not
plausible
Enter correct date
Entered date is not valid.
Enter valid date.
Date not set.
20003
User error: Entered time is not
plausible
Enter correct time
Entered time is not valid.
Enter valid time.
Time not set.
20004
User error: Door was opened
while measurement was
running
Keep door closed during measurement
Door has been opened during
measurement.
Measurement has been cancelled.
Remove cassette and cartridge and close
the door.
20005
User error: Shelf life of test
cassette has expired
Use valid test cassette
Shelf life of the cassette has expired.
Measurement has been cancelled.
Remove cassette and cartridge and close
the door.
20006
Database failure: Test
cassette cannot be processed
Update software
Panel-Type cannot be measured with this
device.
Measurement has been cancelled.
Remove cassette and cartridge and close
the door.
20011
User error: Wrong QA
cassette was used
Ensure correct QA cassette is used
Wrong QA-cassette inserted!
Remove cassette.
20012
User error: QA test was
cancelled
No error
QA-Test cancelled. Counter has not been
reset. Remove cassette. Close door
21200
User error: Printer paper is
missing or printer flap open
Check paper and printer cap
Printer out of paper or printer cover not
closed. Insert paper, close cover and press
OK.
21201
User error: Communication
error (no printer found)
Switch on printer, re-check positioning of
printer. If error persists, service required
Infrared connection between printer
and Analyzer failed.
Switch on printer and align IR port with
Analyzer (place printer on the left).
22000 Service interval expired
Device cannot be used. Maintenance is
required
"Service 1" expired!
Measurement not possible!
Please carry out maintenance.
22000
(QA
Check
failed)
QA Check failed
Device cannot be used. Maintenance is
required.
"QA Check" QA-Test failed!
Measurement not possible!
Please carry out maintenance.
22001 QA Check expired Run QA-Test
"QA-Check" expired!
Measurement not possible!
Run QA-Test(s)!
Error