5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 22
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
2.4 Replacing the Real-Time Clock on the CPU 68332 PCB
Switch the Evita off.
Remove the CPU 68332 PCB.
Remove the real-time clock (D35).
Fig. 18: Component mounting diagram CPU PCB (replacing the real-time clock)
Insert the new real-time clock into the IC socket.
Install the CPU 68332 PCB in the Evita.
Secure the CPU 68332 PCB with the screws.
Switch the Evita on.
Adjust time and date.
Check the Evita by following the instructions in the Evita 4/Evita 2 dura Test Certificate.
1
4
X5 X6 X7
1
2
X4
1
X9
X11
+5 V
GND
+
1
7
X20
X13
X10
C400
+5 V-ISO
GND
D35