5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 68
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
13.2 Replacing the Touchframe
Switch the Evita 4 off.
Open the control unit.
Remove the display.
Remove the four distance pins.
Fig. 36: Touchscreen
Remove the touchscreen.
Mount the touchscreen by following the reverse method used for dismounting.
Check the Evita 4 by following the instructions in the Evita 4 Test Certificate.
IMPORTANT
Dust particles may affect the view on the display. Before mounting the display,
clean the display and the window with a lint-free cloth.
CAUTION
Alcoholic cleaning agents may damage the display and the window. DO NOT
use alcoholic cleaning agents to clean the display or the window.