5664.500/800 Evita 4/Evita 2 dura 06/98 Repair Instructions Page 7
For internal use only. Copyright reserved.
GBR5664500T055664800.fm 18.03.99
Dräger Medizintechnik
D
1.1.1 Sub D Socket Configuration
Important: The voltages can be read out in diagnosis mode refer to section of "Repair
Instructions - Service No./ Modi, Error list" chapter 3.3.2, Diagnosis page “Microprocessor”
of “Electronics”, page 70. or with external DS mode using PC refer to section of "Repair
Instructions - Service No./ Modi, Error list" chapter 9.4.1, Testing voltages, page 138. and
refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 9.4.2, Power
pack status, page 140.
Pin Assignment Pin Assignment
1 +24 V –1680 mV +720 mV 5 +5 V, 15 V ±77.3 mV
2 +15 V ±450 mV 7 Shield
3 AGND 9 +12 V ±240 mV
4 DGND 10 –15 V ±450 mV