11
Operating Concept
Screen Operating Controls
Screen Operating Controls
1 Menu keys for selecting the menu on the screen.
2 Central "turn-and-push" rotary knob for selecting and
setting the options displayed on the screen.
To select/set = turn the rotary knob
To confirm = press the rotary knob.
3 Screen operating keys:
—»Print « key for manual printer logging,
» « key for setting the screen brightness to bright
or dark,
—»Freeze « key for freezing the curves,
—»Curves « key for displaying a different pair of
curves,
—»Values « key for displaying a different
combination of measured values,
one reserve key for future functions.
Keys for routine and additional functions
4 Frequently used keys for routine functions are positioned
on the right-hand side of the front panel:
—»
g« key for suppressing the audible alarm tone,
—»Alarm Reset« key for resetting or acknowledging
alarm messages,
» « key for calling up information and help on the
desired setting,
» « key for protecting against inadvertent or
unauthorised modification of the settings of the
ventilation parameters or ventilation modes.
5 Keys for additional functions are positioned on the left-hand
side of the front panel:
» « key for switching the medicament nebuliser
on/off,
—»O
2 Y Suction« key for oxygen enrichment during
bronchial suction,
—»Insp. hold« for manually activated inspiration,
—»Exp. hold« for extending the expiration time,
one reserve key for future functions.