Technical Data
114
Materials used
Materials used
Cable length Up to 15 m
Load impedance 3000 to 7000 Ω
Signal level (at load impedance 3000 to 7000 Ω)
Low Between 3 and 15 V
High Between –3 and –15 V
Electrical isolation Port COM 1 is electrically isolated from the machine
electronics.
The test voltage for the electrical isolation equals 1500 V.
Pin assignment Pin 2 RxD
Pin 3 TxD
Pin 5 GND
Connector housing Machine housing
Digital output Output for independent lung ventilation (ILV)
Digital output (optional) for output and reception via two RS 232 C interfaces
Digital output (optional) for output and reception via a CAN interface
Analogue output (optional) for output of analog data
Electromagnetic compatibility (EMC)
(conforming to European Directive 89/336/EEC)
Tested in accordance with EN 60601-1-2
Classification
as per EC Directive 93/42/EEC Annex IX
II b
UMDNS-Code
Universal Medical Device Nomenclature System –
Nomenclature for medical products
17-429
Part Appearance Material
Ventilation hose milky, transparent silicone rubber
Water traps yellow, transparent polysulphone
Y-piece
with
connector for temperature measurement
milky, transparent
yellow, transparent
polysulphone
silicone rubber
Expiration valve
housing, closure
white polyamide
Diaphragm whitish and grey silicone rubber and aluminium
CO
2 cuvette yellow, transparent polysulphone with glass windows
Temperature sensor/cable milky/green or blue silicone rubber
CO
2 sensor/cable grey/grey polyurethane