EasyManua.ls Logo

Dräger Evita 2 Dura - Fitting Bacterial Filter; Fitting Humidifier and Ventilation Hoses; For Ventilating Infants

Dräger Evita 2 Dura
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23
Preparing for use
For ventilating infants
For ventilating infants
Up to 300 mL tidal volume VT and below:
Patient mode »Paediatric«
Do not use a heat and moisture exchanger (HME) at the
same time as a humidifier! Risk of increased breathing
resistance because of condensation.
Fitting bacterial filter
Fit the bacterial filter to the inspiratory port.
Fitting humidifier and ventilation hoses
Prepare the "Fisher & Paykel MR 730" humidifier in
accordance with the separate Instructions for Use,
using hose set K (paediatric), as supplied.
1 Hang the humidifier with a bracket to the mounting below
the machine and tighten the screws.
2 Hang the hinged arm with a bracket to the rail on the
left-hand side, and tighten the screws.
Connect the ventilation hoses, and note their length
(in metres).
3 Place the water trap in the vertical position.
Do not place any containers with liquid on or above the
ventilator! Any leaking or spilled liquid can cause
malfunctions!
062
D
Evita 2 dura
063
0.6m
0.6m
0.4m
1.1m
1
2
3