23
Preparing for use
For ventilating infants
For ventilating infants
Up to 300 mL tidal volume VT and below:
Patient mode »Paediatric«
Do not use a heat and moisture exchanger (HME) at the
same time as a humidifier! Risk of increased breathing
resistance because of condensation.
Fitting bacterial filter
● Fit the bacterial filter to the inspiratory port.
Fitting humidifier and ventilation hoses
● Prepare the "Fisher & Paykel MR 730" humidifier in
accordance with the separate Instructions for Use,
using hose set K (paediatric), as supplied.
1 Hang the humidifier with a bracket to the mounting below
the machine and tighten the screws.
2 Hang the hinged arm with a bracket to the rail on the
left-hand side, and tighten the screws.
● Connect the ventilation hoses, and note their length
(in metres).
3 Place the water trap in the vertical position.
Do not place any containers with liquid on or above the
ventilator! Any leaking or spilled liquid can cause
malfunctions!
062
D
Evita 2 dura
063
0.6m
0.6m
0.4m
1.1m
1
2
3