Preparing for use
24
In-house/intra hospital transport
If no instrument tray (option) is fitted to the device
In-house/intra hospital transport
Ensure that the equipment does not topple over by moving the
accessories as close to the ventilator and trolley as possible:
● Hinged arm set to minimum deflection.
● Drawers pushed in fully.
● Hoses and cables hooked as close as possible to the
trolley.
● Humidifier secured to the trolley, not to the ceiling pendant/
bed/wall.
If no instrument tray (option) is fitted to the
device
Seal off the slits in the top panel with the rubber plugs:
1 Press the round plugs in the front slits – rounded part
facing outwards.
2 Press the flat plugs into the rear slits.
Do not place any liquid container (e.g. infusion container)
above or on top of the Evita 2 dura!
Any leak, spill or seepage could prevent it working properly.
The unit must not be tilted more than 5
o
when mounted
on a trolley.
It may tip over at steeper angles.
066
D
Evita 2 dura
11