Preparing for use
26
Supply and Connections
In the case of a power interruption lasting longer then
10 milliseconds, the machine restarts with a short self-test
lasting about 4 seconds – ventilation is continued with the
same values that were set before the power interruption.
If a lower alarm limit has been set for the minute volume, the
»!!! MV low« alarm is activated until the measured value has
risen above the lower alarm limit.
With optional Evita DC power supply:
Follow Instructions for Use of the Evita DC power supply.
Gas supply
Screw the connecting hoses for medical air and oxygen to
the back panel of Evita 2 dura and insert their probes into
the terminal units.
The compressed gases must be dry and free from dust and
oil. Gas pressure must be 3 to 6 bar.
Other equipment, e.g. printers, may only be connected to
the COM ports if Evita 2 dura is connected to the mains
power supply via a mains power cable or if it has been
earthed via the earth connection on the back of the unit.
Electric power may pose a hazard in all other cases.