3
What's new in Evita 2 dura
software 4.n
Specification of the humidifier used
— »Active humidifier«
or
— »HME/Filter« (artificial nose)
— for more accurate measurement of the volume parameters
Apnoea ventilation On/Off
— can be selected as starting configuration
Extended range of settings for the alarm time T
Apnoea
>
— from 5 to 60 seconds
(formerly 15 to 60 seconds)
Ventilation mode BIPAP
Assist
— for pressure-controlled assisted ventilation
Patient mode »prev. patient« can be selected
— to adopt the settings, including alarms, which were
effective before switching off the equipment
Leakage compensation On/Off
— for activation and deactivation of the automatic leakage
compensation function
Monitoring of tube blockages
— New alarm message »Tube blocked !!!«
Additional weaning parameters
available as software version 4.n plus upgrade
in addition to the parameter occlusion pressure P 0.1
Evita 2 dura 4.n also determines the parameters
— RSB Rapid Shallow Breathing index
and
— NIF Negative Inspiratory Force Index
—f
spn
and MV
spn
as trend
External flow source
available as software version 4.n plus upgrade
— The amount of external flow is calculated by Evita 2 dura 4.n
(e.g. for additional tracheal gas insufflation) and adjusts the
volume monitoring tolerances in order to avoid inadvertent
alarms
Evita Remote (Remote Pad)
optionally available
— Remote control pad for parallel remote operation of
function keys on Evita 2 dura 4.n
NIV
optionally available
— Application mode to support non-invasive ventilation
therapies
Nurse call
optionally available
— Socket for connecting alarm signals to a central alarm
station in the hospital
Simplified settings
under »other modes«