Configuration
78
Ventilation Defaults
Default ventilation parameter values
These defaults specify the ventilation parameters and alarm
limits that are not patient-specific and are activated on
starting up the device.
Press the »Calib./Config.« menu key.
Press the »Ventilation jj« menu key.
Enter code number »3032«.
Select the »Parameter« menu with the »Ventilation jj«
menu key.
Display (example):
Select the Pmax desired screen field = turn the rotary knob.
Confirm = press the rotary knob.
Set value = turn the rotary knob.
Confirm = press the rotary knob.
Default ventilation parameters
The hospital-specific defaults for your hospital can be entered
in the table.
To restore the factory-set defaults:
Select the »Basic settings« screen field = turn the rotary
knob.
Confirm = press the rotary knob.
I
: EPinsp
mbar
P
ASB
mbar
PEEP
mbar
Ramp
s
Trigger
L/min
O
2
Vol.%
Factory setting 1:2 15 0 5 0.2 5 30
Hospital-
specific
setting
526