79
Configuration
Ventilation Defaults
Activating/deactivating pressure limit Pmax
— This parameter defines the pressure limit for pressure-
limited ventilation in the IPPV, SIMV and MMV ventilation
modes.
● Press the »Calib./Config.« menu key.
● Press the »Ventilation jj« menu key.
● Enter code number »3032«.
● With the »Ventilation jj« menu key, select the
»Parameter« menu.
Display (example):
● Select the »Pmax« screen key = turn the rotary knob.
The "LED" in the screen key »Pmax« lights up yellow.
Pmax is activated.
● Deactivate Pmax = press rotary knob, the "LED" in the
screen key »Pmax« lights up green.
Set the value for Pmax:
● Press the »Pinsp« parameter key.
● Set the desired value = turn the rotary knob.
Confirm = press the rotary knob.
534040
D
Evita 2 dura